Skip to content
soundlevelmeter.review

Norsonic Nor140 vs Nova

Every specification Norsonic and Pulsar Instruments publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationNorsonic Nor140NorsonicNovaPulsar Instruments
StatusCurrentCurrent
Verified on
Manufacturer list priceGBP 1,100
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1 class 1, IEC 61260-1 class 1, ANSI S1.4, ANSI S1.11, DIN 45657, IEC 60268-16 ed. 5 (STIPA, option 13), ISO 16283-1/-2/-3, ISO 717-1/-2, ISO 3382-2, ISO 1996-1/-2IEC 61672-1:2013 class 1 or class 2, group X, IEC 61672-2:2013, IEC 60651:2001 type 1 I or type 2 I, IEC 60804:2000 type 1 or type 2, IEC 61252:1993, ANSI S1.4-1983 (R2006), ANSI S1.43-1997 (R2007), ANSI S1.25:1991, IEC 61260 and ANSI S1.11-2004 (1/1 octave filters)
Pattern approvals (differs)PTB (Germany), METAS (Switzerland), LNE 24006 rev. 2 (France), BEV Z 1522-08 (Austria)—
Recommended calibration interval (differs)—12 months
Measurement
Channels11
Measurement range (differs)about -10 to 137 dB at 50 mV/Pa, single 120 dB range (147 dB with option 18)20 to 140 dB RMS, single range
Peak range (differs)140 dB peak standard; 150 dB with option 18up to 143 dB
Frequency range (differs)0.5 Hz to 16 kHz (1/1 octave, option 1); 0.4 Hz to 20 kHz (1/3 octave, option 3)—
Frequency weightingsA, C, ZA, C, Z
Time weightingsF, S, IF, S, I
Simultaneous weightings (differs)NoYes
Parameters (differs)SPL, Leq, Lmin, Lmax, LE, Lpeak, LN (8 values), LeqI, LEI, LTmax5Lxy, Lxy max/min, Lxeq, LCpeak, LZpeak, LApeak, LCeq-LAeq, LxE, LAIeq, Ln (14 values), TWA, Dose %, estimated Dose %
Dosimetry (differs)—Yes
Exchange rates (differs)—3 dB, 4 dB, 5 dB
Analysis
Octave bands (differs)1/1 and 1/31/1
Octave band filter class (differs)class 1 (IEC 61260), options 1 and 3IEC 61260 and ANSI S1.11-2004, 31.5 Hz to 16 kHz (Models 45/46)
FFT (differs)8000 lines, 1.46 Hz resolution (option 14)—
Reverberation time (differs)YesNo
Building acoustics (differs)YesNo
Speech intelligibility (differs)STIPA (option 13)—
Audio recordingYesYes
Logging and data
Shortest logging interval (differs)5 ms (option 7); 1 s standard (option 6)10 ms (time history); periodic 1 min to 1 h and Lden
Memory (differs)About 25 MB internal flash plus SD card (included)4 GB removable memory card
Hardware
Microphone (differs)1/2" free-field Nor1225, 50 mV/Pa, with Nor1209 preamplifier (7-pin LEMO); 0 or 200 V polarisation1/2" prepolarised PM1 (class 1) or PM2 (class 2); PA40 removable preamplifier
Detachable preamplifierYesYes
Outdoor microphone kit (differs)Nor1217 outdoor microphone protectionWK3-N outdoor kit with PM172 outdoor microphone
Display (differs)Monochrome transflective backlit LCD, 160 x 240OLED with ambient light sensor (product description says TFT)
Battery (differs)4 x AA (LR6); 11 to 16 V DC external6 x AA; external 5 to 15 V DC via I/O socket
Battery life (differs)8 h30 h
Weight (differs)410 g440 g
Dimensions (mm) (differs)210 x 75 x 30 (292 with microphone and preamplifier)280 x 80 x 42 (with preamplifier and microphone)
Operating temperature (°C)-10 to 50-10 to 50
Connectivity and software
Connectivity (differs)USB 2.0, RS-232, 3 digital inputs, 4 digital outputs, AC output (3.5 mm), Bluetooth via Nor520 transceiverUSB 2.0 Mini-B, 3.5 mm AC output, multipin I/O (AC and DC output)
Remote control (differs)NorCloud login—
Software (differs)Nor850 Suite, NorCloud, CtrlBuild, NorBuild (Nor1028)AnalyzerPlus
Latest firmware (differs)4.0.1550 (listed 14 March 2024)—

The instruments on their own