Compare instruments
Up to 4 instruments of the same type, every published specification side by side. Rows where the values differ are tinted. A blank cell means the manufacturer does not publish that value.
Comparing 2 noise dosimeters. Back to the category table.
| Specification | dBadge CEL-35xCasella | Rion NB-14Rion |
|---|---|---|
| Status | Discontinued | Current |
| Verified on | ||
| Standards and class | ||
| IEC 61672-1 class (differs) | — | Class 2 |
| Standards (differs) | IEC 61252:2002, BS EN 61252:1997, ANSI S1.25 | IEC 61252:1993+AMD1:2000+AMD2:2017, IEC 61672-1:2013 class 2, JIS C 1509-1:2017 class 2 |
| Measurement | ||
| Channels | 1 | 1 |
| Measurement range (differs) | 65.0 to 140.3 dB(A) RMS | 58 to 143 dB (LAeq, LCeq) |
| Linearity range (differs) | 65.0 to 140.3 dB(A) | — |
| Peak range (differs) | 95.0 to 143.3 dB (C or Z) | 75 to 146 dB (LCpeak) |
| Frequency range (differs) | — | 20 Hz to 8 kHz |
| Frequency weightings (differs) | A, C, Z | A, C |
| Time weightings (differs) | F, S, I | — |
| Simultaneous weightings (differs) | — | Yes |
| Parameters (differs) | LAeq, LCeq (Plus), Lavg, LCpeak, LZpeak, LAE, LASmax, TWA, LEP,d, Pa2h | LAeq, LCeq, LCpeak, Dose |
| Dosimetry | Yes | Yes |
| Exchange rates (differs) | 3 dB, 5 dB | — |
| Intrinsically safe marking (differs) | ATEX I M1, II 1G, EEx ia IIC T2 (222 °C), EEx ia I; Ta -20 to +40 °C; microphone FM/CSA Class 1 Division 1 Groups A-D T2 (IS models) | — |
| Analysis | ||
| Octave bands | None | None |
| Reverberation time | No | No |
| Building acoustics | No | No |
| Audio recording | No | No |
| Logging and data | ||
| Shortest logging interval (differs) | 1 min | 1 s |
| Memory (differs) | — | Internal, about 48 h of data |
| Hardware | ||
| Microphone (differs) | 1/2" CEL-252, 30 mV/Pa, type 2 | UC-52 |
| Detachable preamplifier | No | No |
| Display (differs) | LCD with blue/red alarm LED | Status indicators and dose bar graph |
| Battery (differs) | internal NiMH; full charge under 90 min | Built-in rechargeable lithium-ion, charged via USB cradle |
| Battery life (differs) | 28 h | 12 h |
| Weight (differs) | 68 g | 72 g |
| Dimensions (mm) (differs) | 72 x 47 x 52 | 85 x 43 x 22 |
| IP rating (differs) | — | IP54 excluding microphone |
| Operating temperature (°C) (differs) | 0 to 40 | -10 to 50 |
| Connectivity and software | ||
| Connectivity (differs) | infrared (193200B cable) | USB via cradle |
| Remote control (differs) | — | Mobile app for viewing results (Google Play; name not given) |
| Software (differs) | Casella Insight, dB35 | AS-05 Viewer Noise Dosimeter Data Management Software |
| Latest firmware (differs) | 193026-01-15 (firmware file on support page) | — |