Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationdBadge2CasellaNoiseBadge (Model 22)Pulsar Instruments
StatusCurrentCurrent
Verified on
Manufacturer list priceGBP 510
Standards and class
Standards (differs)IEC 61252 Ed. 1.1b:2002, ANSI S1.25-1991 (R2007), IEC 61260:2014 class 2 (Pro filters), ANSI/ASA S1.11-2014 class 2 (Pro filters)IEC 61252:1993, ANSI S1.25:1991, IEC 60942:2001 class 2 (reader calibrator)
Recommended calibration interval (differs)—12 months
Measurement
Channels (differs)12
Measurement range (differs)54.0 to 140.3 dB(A) RMS, dynamic range 96 dB70 to 130 dB(A) RMS
Linearity range (differs)54.0 to 140.3 dB(A)—
Peak range (differs)90.0 to 143.3 dB (C or Z)120 to 140 dB(C)
Frequency weightings (differs)A, C, ZA, C
Time weightings (differs)F, S, IS
Parameters (differs)Lavg, TWA(8h), projected TWA, Dose, projected Dose, LAeq, LCeq, LAIeq, LEP,d / LEX,8h, LApk, LCpk, LZpk, LAFmax, LASmax, LAImax, Pa2h, LC-LA (HML), Lpk counts > 135/137/140 dB, LAE, Motion IndexLAeq, LEP,d / LEX,8h, LAE, LCpeak, Lavg, TWA, Dose %, estimated Dose %, exposure Pa2h, 115 dB(A) exceedance
DosimetryYesYes
Exchange rates (differs)3 dB, 5 dB3 dB, 4 dB, 5 dB
Intrinsically safe marking (differs)ATEX/IECEx: I M1; II 1GD; Ex ia I Ma; Ex ia IIC T4 (135 °C) Ga; Ex ia IIIC T135 °C Da; Ta = -20 °C to +45 °C (IS models)—
Analysis
Octave bands (differs)1/1None
Octave band filter class (differs)class 2 (IEC 61260:2014), 9 bands 31.5 Hz to 8 kHz (Pro)—
Reverberation timeNoNo
Building acousticsNoNo
Audio recording (differs)YesNo
Logging and data
Shortest logging interval (differs)1 s1 min
Memory (differs)internal microSD, 300 h of run data including 1 s profile; up to 90 min event audioreader stores up to 93 eight-hour, 64 twelve-hour or 33 twenty-four-hour measurements
Hardware
Microphone (differs)1/2" CEL-252, 31.6 mV/Pa, type 2removable prepolarised electret, 1/2" class 2
Detachable preamplifierNoNo
Display (differs)96 x 64 full-colour OLEDnone on badge (status LED); backlit LCD on reader
Battery (differs)internal NiMH 800 mAh; about 10 h charge (manual also states about 6 h)internal NiMH, about 3 h charge; reader 2 x AA or USB
Battery life (differs)55 h20 h
Weight (differs)117 g45 g
Dimensions (mm) (differs)85 x 54 x 55height 38, base diameter 47, microphone apex diameter 13
IP rating (differs)IP65 (with permeable air vent)—
Operating temperature (°C) (differs)0 to 40 (for IEC 61252 compliance); -10 to 50 within 0.8 dB-10 to 50
Connectivity and software
Connectivity (differs)Bluetooth 4.0 Low Energyinfrared to reader, USB 2.0 (reader to PC)
Remote control (differs)Airwave app (Android)—
Software (differs)Casella Insight, AirwaveAnalyzerPlus