SpecificationdBadge2CasellaTES-660TES Electrical Electronic
StatusCurrentCurrent
Verified on
Standards and class
Standards (differs)IEC 61252 Ed. 1.1b:2002, ANSI S1.25-1991 (R2007), IEC 61260:2014 class 2 (Pro filters), ANSI/ASA S1.11-2014 class 2 (Pro filters)IEC 61252:2017, ANSI S1.25:1991
Measurement
Channels11
Measurement range (differs)54.0 to 140.3 dB(A) RMS, dynamic range 96 dB70 to 140 dB
Linearity range (differs)54.0 to 140.3 dB(A)—
Peak range (differs)90.0 to 143.3 dB (C or Z)103 to 143 dB
Frequency weightingsA, C, ZA, C, Z
Time weightingsF, S, IF, S, I
Parameters (differs)Lavg, TWA(8h), projected TWA, Dose, projected Dose, LAeq, LCeq, LAIeq, LEP,d / LEX,8h, LApk, LCpk, LZpk, LAFmax, LASmax, LAImax, Pa2h, LC-LA (HML), Lpk counts > 135/137/140 dB, LAE, Motion IndexLavg, LZpk, Leq, LCpk
DosimetryYesYes
Exchange rates (differs)3 dB, 5 dB3 dB, 4 dB, 5 dB, 6 dB
Intrinsically safe marking (differs)ATEX/IECEx: I M1; II 1GD; Ex ia I Ma; Ex ia IIC T4 (135 °C) Ga; Ex ia IIIC T135 °C Da; Ta = -20 °C to +45 °C (IS models)TES-660A only: ATEX, FM/CSA; EN 60079-0:2006, EN 60079-11:2007, EN 60079-26:2007
Analysis
Octave bands (differs)1/1—
Octave band filter class (differs)class 2 (IEC 61260:2014), 9 bands 31.5 Hz to 8 kHz (Pro)—
Reverberation time (differs)No—
Building acoustics (differs)No—
Audio recording (differs)Yes—
Logging and data
Shortest logging interval (differs)1 s—
Memory (differs)internal microSD, 300 h of run data including 1 s profile; up to 90 min event audio30 measurements
Hardware
Microphone (differs)1/2" CEL-252, 31.6 mV/Pa, type 21/2" electret, field-replaceable, threaded
Detachable preamplifier (differs)No—
Display (differs)96 x 64 full-colour OLED128 x 64 graphic
Battery (differs)internal NiMH 800 mAh; about 10 h charge (manual also states about 6 h)Rechargeable Li-polymer 3.7 V 470 mAh
Battery life (differs)55 h—
Weight (differs)117 g77 g
Dimensions (mm) (differs)85 x 54 x 5584 x 49 x 55 (with windscreen)
IP rating (differs)IP65 (with permeable air vent)—
Operating temperature (°C) (differs)0 to 40 (for IEC 61252 compliance); -10 to 50 within 0.8 dB0 to 50
Connectivity and software
Connectivity (differs)Bluetooth 4.0 Low EnergyUSB via DS-01 charging unit
Remote control (differs)Airwave app (Android)—
Software (differs)Casella Insight, Airwave—

Change the selection