Skip to content
soundlevelmeter.review

Compare instruments

Up to 4 instruments of the same type, every published specification side by side. Rows where the values differ are tinted. A blank cell means the manufacturer does not publish that value.

Comparing 2 handheld sound level meters. Back to the category table.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationCastle dBAirCastle GroupXPT800Delta OHM (Senseca)
StatusCurrentCurrent
Verified on
Manufacturer list priceGBP 2,675
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2013, IEC 61260-1:2014 (where octave bands fitted), IEC 61252:1993 amendment 1:2000 (where exposure fitted)IEC 61672-1:2013 class 1, IEC 60651 and IEC 60804 type 1, IEC 61260-1:2014 (1/1 and 1/3 octave filters), ANSI/ASA S1.4-2014 class 1, ANSI S1.4-1983 (S1.4A-1985) type 1, ANSI S1.43-1997 type 1, ANSI/ASA S1.11-2014 Part 1, ISO 1996-2 and ISO/PAS 1996-3:2022 (tonality and impulsiveness)
Measurement
Channels11
Measurement range (differs)Linear operating range 95 dBA: 20 to 137 dB; C: 22 to 137 dB; Z: 25 to 137 dB (linear, with MC800); 15 dBA to 140 dB peak quoted as 123 dB linearity range
Linearity range (differs)—123 dB (15 dBA to 140 dB peak); dynamic range greater than 125 dB
Peak range (differs)—up to 140 dB peak (MC800); up to 178 dB with MS802 set
Frequency range (differs)12.5 Hz to 20 kHz including class 1 microphone (1 Hz to 20 kHz electrical)3.15 Hz to 20 kHz (microphone); 1/1 octave 8 Hz to 16 kHz; 1/3 octave 6.3 Hz to 20 kHz
Frequency weightings (differs)A, C, ZA, B, C, Z
Time weightingsF, S, IF, S, I
Simultaneous weightingsYesYes
Parameters (differs)LSPL, LE, Leq, Lmax, Lmin, Peak, Ltm3, Ltm5, Lday, Lnight, Ldn, Lden, NA, Ln (ten user-definable), LAF, LEP,d, Exposure points, Dose, Hearing protector calculatorLp, Leq, LIeq, SEL, Leq moving, Lmax, Lmin, Lpeak, LCeq-LAeq, LUeq, LAFT / LAFTeq, Lden, Ldn, Lday, Levening, Lnight, LN (0.1 to 99.9 %), NC, RNCm, NR, RC (option OF9)
Dosimetry (differs)Yes—
Analysis
Octave bands1/1 and 1/31/1 and 1/3
Octave band filter class (differs)—IEC 61260-1:2014; class not stated on datasheet
FFT (differs)—up to 16,000 lines, simultaneous with octave spectra at 100 ms (datasheet marks FFT specifications as to be announced)
Reverberation time (differs)—Yes
Building acoustics (differs)—Yes
Speech intelligibility (differs)—STI and STIPA (option OF6)
Audio recordingYesYes
Logging and data
Shortest logging interval10 ms10 ms
Memory (differs)16, 32 or 64 GB internal4 GB internal eMMC; USB flash drive; microSD up to 64 GB (to be announced)
Hardware
Microphone (differs)1/2" pre-polarised condenser, 50 mV/Pa, class 1 (Safety version: 25 mV/Pa class 2 microphone selectable)MC800 1/2" free-field, 50 mV/Pa, IEC 61094-4 WS2F, with MP800 preamplifier
Detachable preamplifier (differs)—Yes
Outdoor microphone kit (differs)—HDWME outdoor microphone protection (windscreen, rain shield, bird spike)
Display (differs)2.4 inch colour TFT, 240 x 3204.3" colour touch TFT, 480 x 800, sunlight readable
Battery (differs)4 x AA (1.5 V)Li-ion polymer 9000 mAh rechargeable
Battery life (differs)12 h24 h
Weight (differs)480 g505 g
Dimensions (mm) (differs)210 x 70 x 30 (with preamplifier); 145 x 70 x 30 (without)304 x 86 x 38
Connectivity and software
Connectivity (differs)Micro USB, USB, WiFiUSB-C OTG 2.0, WiFi IEEE 802.11 b/g/n, Ethernet 10/100 (RJ45), LTE/4G (optional OH3H module), RS-232/RS-485 (optional OH3A module), 3.5 mm audio and trigger I/O, RJ12 auxiliary power
Remote control (differs)Castle CloudNS-Storage cloud with push sync over 4G or Wi-Fi; NS-Monitor web (planned)
Software (differs)Castle Cloud (3-year Pro licence included)NS-ENS Environmental Noise Studio, NS-SIS Sound Insulation Studio, NS-Storage, NoiseStudio 2.0

Change the selection

Choose up to 4 instruments of the same type. Instruments are grouped by brand; the type is shown after the model name.

Compare sound level meters | soundlevelmeter.review