Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationdoseBadge5 Personal Noise Dosimeter (CR:120A)Cirrus ResearchTES-660TES Electrical Electronic
StatusCurrentCurrent
Verified on
Standards and class
Standards (differs)IEC 61252:1993 +AMD1:2000 (Ed 1.1, 2002-03), ANSI S1.25:1991 (R2017), IEC 61260-1:2014 class 2 and ANSI S1.11-2014 class 2 (octave bands)IEC 61252:2017, ANSI S1.25:1991
Measurement
Channels (differs)61
Measurement range (differs)60 to 140 dB(A) RMS; linear operating range 65 to 140 dB(A)70 to 140 dB
Linearity range (differs)65 to 140 dB(A)—
Peak range (differs)80 to 143 dB(C)103 to 143 dB
Frequency range (differs)63 Hz to 8 kHz (octave bands)—
Frequency weightingsA, C, ZA, C, Z
Time weightings (differs)F, SF, S, I
Simultaneous weightings (differs)Yes—
Parameters (differs)Leq/LAVG, LEP,d / LEX,8h / TWA, %Dose and estimated %Dose, Lpeak, sound exposure, ULT duration, SPL max and min, LCeq-LAeq (HML)Lavg, LZpk, Leq, LCpk
DosimetryYesYes
Exchange rates (differs)3 dB, 4 dB, 5 dB3 dB, 4 dB, 5 dB, 6 dB
Intrinsically safe marking (differs)—TES-660A only: ATEX, FM/CSA; EN 60079-0:2006, EN 60079-11:2007, EN 60079-26:2007
Analysis
Octave bands (differs)1/1—
Octave band filter class (differs)class 2 (IEC 61260-1:2014)—
Logging and data
Shortest logging interval (differs)1 s—
Memory (differs)up to 80 h of time history (6 channels); up to 40 measurements; 24 h maximum single measurement30 measurements
Hardware
Microphone (differs)—1/2" electret, field-replaceable, threaded
Detachable preamplifier (differs)No—
Display (differs)none128 x 64 graphic
Battery (differs)internal NiMHRechargeable Li-polymer 3.7 V 470 mAh
Battery life (differs)20 h—
External power (differs)charging dock (3 h from empty)—
Weight (differs)85 g77 g
Dimensions (mm) (differs)66 x 43 x 53 (excluding clips)84 x 49 x 55 (with windscreen)
Operating temperature (°C) (differs)-10 to 500 to 50
Connectivity and software
Connectivity (differs)Bluetooth, USB via charging dockUSB via DS-01 charging unit
Remote control (differs)doseBadge5 remote control wand; dBLink / MyCirrus app (iOS, Android)—
Software (differs)NoiseTools, dBLink app—

Change the selection

Choose up to 4 instruments of the same type. Instruments are grouped by brand; the type is shown after the model name.

Compare sound level meters | soundlevelmeter.review