SpecificationdoseBadge5 Personal Noise Dosimeter (CR:120A)Cirrus ResearchQuest Edge 4+ (EG4P)TSI Quest
StatusCurrentCurrent
Verified on
Standards and class
Standards (differs)IEC 61252:1993 +AMD1:2000 (Ed 1.1, 2002-03), ANSI S1.25:1991 (R2017), IEC 61260-1:2014 class 2 and ANSI S1.11-2014 class 2 (octave bands)ANSI S1.25 (personal noise dosimeters), IEC 61252 (personal sound exposure meters)
Measurement
Channels (differs)61
Measurement range (differs)60 to 140 dB(A) RMS; linear operating range 65 to 140 dB(A)70 to 140 dB
Linearity range (differs)65 to 140 dB(A)—
Peak range (differs)80 to 143 dB(C)110 to 143 dB
Frequency range (differs)63 Hz to 8 kHz (octave bands)—
Frequency weightingsA, C, ZA, C, Z
Time weightingsF, SF, S
Simultaneous weightings (differs)Yes—
Parameters (differs)Leq/LAVG, LEP,d / LEX,8h / TWA, %Dose and estimated %Dose, Lpeak, sound exposure, ULT duration, SPL max and min, LCeq-LAeq (HML)Lavg, Leq, SPL, Lmax, Lmin, SEL (LEP), Dose, TWA, Lpeak, exposure, time above threshold, upper limit time
DosimetryYesYes
Exchange rates3 dB, 4 dB, 5 dB3 dB, 4 dB, 5 dB
Analysis
Octave bands (differs)1/1None
Octave band filter class (differs)class 2 (IEC 61260-1:2014)—
Logging and data
Shortest logging interval (differs)1 s1 min
Memory (differs)up to 80 h of time history (6 channels); up to 40 measurements; 24 h maximum single measurement180 hours
Hardware
Microphone (differs)—1/4" prepolarised condenser (electret), threaded, field-replaceable
Detachable preamplifier (differs)No—
Display (differs)none128 x 64 colour OLED with configurable auto shut-off
Battery (differs)internal NiMHLithium polymer flat cell; charged in EdgeDock (2 to 4 h)
Battery life20 h20 h
External power (differs)charging dock (3 h from empty)—
Weight85 g85 g
Dimensions (mm) (differs)66 x 43 x 53 (excluding clips)88 x 53 x 19
IP rating (differs)—IP65
Operating temperature (°C) (differs)-10 to 50-10 to 50; storage -25 to 60
Connectivity and software
Connectivity (differs)Bluetooth, USB via charging dockBluetooth (optional), USB via docking station
Remote control (differs)doseBadge5 remote control wand; dBLink / MyCirrus app (iOS, Android)Edge dB mobile app (iOS, Android)
Software (differs)NoiseTools, dBLink appDetection Management Software (DMS), Edge dB app