Compare instruments
Up to 4 instruments of the same type, every published specification side by side. Rows where the values differ are tinted. A blank cell means the manufacturer does not publish that value.
Comparing 2 handheld sound level meters. Back to the category table.
| Specification | Model 831Larson Davis | NovaPulsar Instruments |
|---|---|---|
| Status | Discontinued | Current |
| Verified on | ||
| Manufacturer list price | GBP 1,100 | |
| Standards and class | ||
| IEC 61672-1 class | Class 1 | Class 1 |
| Standards (differs) | IEC 61672-1 class 1 (manual states measuring to IEC 61672-1 and class 1 performance; full standards list in Appendix A not read) | IEC 61672-1:2013 class 1 or class 2, group X, IEC 61672-2:2013, IEC 60651:2001 type 1 I or type 2 I, IEC 60804:2000 type 1 or type 2, IEC 61252:1993, ANSI S1.4-1983 (R2006), ANSI S1.43-1997 (R2007), ANSI S1.25:1991, IEC 61260 and ANSI S1.11-2004 (1/1 octave filters) |
| Recommended calibration interval (differs) | — | 12 months |
| Measurement | ||
| Channels | 1 | 1 |
| Measurement range (differs) | 16 to 140 dB with PRM831 preamplifier and 377B02 microphone (dynamic range > 120 dB(A)) | 20 to 140 dB RMS, single range |
| Peak range (differs) | — | up to 143 dB |
| Frequency weightings | A, C, Z | A, C, Z |
| Time weightings | F, S, I | F, S, I |
| Simultaneous weightings (differs) | — | Yes |
| Parameters (differs) | SPL, Leq, Lmax, Lmin, Lpeak, Lpeak(max), LN statistics, sound exposure and sound exposure level, Ldn, Lden, CNEL, noise dose (831-IH), RT60 (831-RT) | Lxy, Lxy max/min, Lxeq, LCpeak, LZpeak, LApeak, LCeq-LAeq, LxE, LAIeq, Ln (14 values), TWA, Dose %, estimated Dose % |
| Dosimetry | Yes | Yes |
| Exchange rates (differs) | — | 3 dB, 4 dB, 5 dB |
| Analysis | ||
| Octave bands (differs) | 1/1 and 1/3 | 1/1 |
| Octave band filter class (differs) | 831-OB3 option (class not stated in manual text read) | IEC 61260 and ANSI S1.11-2004, 31.5 Hz to 16 kHz (Models 45/46) |
| FFT (differs) | 831-FFT option (line count not stated in manual text read) | — |
| Reverberation time (differs) | Yes | No |
| Building acoustics (differs) | — | No |
| Audio recording | Yes | Yes |
| Logging and data | ||
| Shortest logging interval (differs) | 20 ms (831-LOG); 2.5 ms spectral logging (831-FST) | 10 ms (time history); periodic 1 min to 1 h and Lden |
| Memory (differs) | 2 GB internal data storage | 4 GB removable memory card |
| Hardware | ||
| Microphone (differs) | 1/2" 377B02 free-field or 377C20 random incidence, 50 mV/Pa, class 1 | 1/2" prepolarised PM1 (class 1) or PM2 (class 2); PA40 removable preamplifier |
| Detachable preamplifier | Yes | Yes |
| Outdoor microphone kit (differs) | — | WK3-N outdoor kit with PM172 outdoor microphone |
| Weather sensors (differs) | Yes | — |
| Display (differs) | 160 x 240 graphic LCD with backlight | OLED with ambient light sensor (product description says TFT) |
| Battery (differs) | 4 x AA (NiMH, alkaline or 1.5 V lithium) | 6 x AA; external 5 to 15 V DC via I/O socket |
| Battery life (differs) | 8 h | 30 h |
| External power (differs) | +10.8 to +30 V DC; PSA029 adapter 90 to 274 V AC, 5 V DC output | — |
| Weight (differs) | — | 440 g |
| Dimensions (mm) (differs) | — | 280 x 80 x 42 (with preamplifier and microphone) |
| IP rating (differs) | IP53 (case, dust tight) | — |
| Operating temperature (°C) (differs) | — | -10 to 50 |
| Connectivity and software | ||
| Connectivity (differs) | USB 2.0 host and peripheral, AUX connector, I/O connector | USB 2.0 Mini-B, 3.5 mm AC output, multipin I/O (AC and DC output) |
| Software (differs) | G4 LD Utility, SLM Utility-G3 | AnalyzerPlus |
| Latest firmware (differs) | no new firmware from September 2025 (manufacturer announcement); manual I831.01 Rev T (2022) is the last edition | — |