Skip to content
soundlevelmeter.review

Compare instruments

Up to 4 instruments of the same type, every published specification side by side. Rows where the values differ are tinted. A blank cell means the manufacturer does not publish that value.

Comparing 2 handheld sound level meters. Back to the category table.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationNorsonic Nor131NorsonicNovaPulsar Instruments
StatusDiscontinuedCurrent
Verified on
Manufacturer list priceGBP 1,100
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1 class 1 (Nor131) / class 2 (Nor132), IEC 61260-1/-2/-3, ANSI S1.4, ANSI S1.11, ANSI S1.43, DIN 45657, IEC 60268-16 ed. 4 and ed. 5 (STIPA, option 5)IEC 61672-1:2013 class 1 or class 2, group X, IEC 61672-2:2013, IEC 60651:2001 type 1 I or type 2 I, IEC 60804:2000 type 1 or type 2, IEC 61252:1993, ANSI S1.4-1983 (R2006), ANSI S1.43-1997 (R2007), ANSI S1.25:1991, IEC 61260 and ANSI S1.11-2004 (1/1 octave filters)
Recommended calibration interval (differs)—12 months
Measurement
Channels11
Measurement range (differs)20 to 140 dB in a single span; one range covering 120 dB without range changing20 to 140 dB RMS, single range
Peak range (differs)140 dB C-weighted peak; maximum RMS 137 dB(A)up to 143 dB
Frequency range (differs)20 Hz to 16 kHz broadband; filters 8 Hz to 16 kHz (1/1) or 6.3 Hz to 20 kHz (1/3)—
Frequency weightingsA, C, ZA, C, Z
Time weightingsF, S, IF, S, I
Simultaneous weightingsYesYes
Parameters (differs)SPL, Leq, Lmax, Lmin, LE, Lpeak, Tmax5, L1, L5, L10, L50, L90, L95, L99 plus one user percentileLxy, Lxy max/min, Lxeq, LCpeak, LZpeak, LApeak, LCeq-LAeq, LxE, LAIeq, Ln (14 values), TWA, Dose %, estimated Dose %
Dosimetry (differs)—Yes
Exchange rates (differs)—3 dB, 4 dB, 5 dB
Analysis
Octave bands (differs)1/1 and 1/31/1
Octave band filter class (differs)1/1 octave option 1, 12 bands 8 Hz to 16 kHz; 1/3 octave option 4, 36 bands 6.3 Hz to 20 kHzIEC 61260 and ANSI S1.11-2004, 31.5 Hz to 16 kHz (Models 45/46)
Reverberation time (differs)—No
Building acoustics (differs)—No
Speech intelligibility (differs)STIPA (option 5)—
Audio recording (differs)—Yes
Logging and data
Shortest logging interval (differs)—10 ms (time history); periodic 1 min to 1 h and Lden
Memory (differs)5 MB internal non-volatile; about 10,000 individual measurements4 GB removable memory card
Hardware
Microphone (differs)1/2" Nor1228 with Nor1207 IEPE preamplifier1/2" prepolarised PM1 (class 1) or PM2 (class 2); PA40 removable preamplifier
Detachable preamplifierYesYes
Outdoor microphone kit (differs)Nor1218 outdoor microphone protectionWK3-N outdoor kit with PM172 outdoor microphone
Display (differs)Monochrome transflective LCD, 160 x 240, backlitOLED with ambient light sensor (product description says TFT)
Battery (differs)4 x AA (LR6) alkaline; 11 to 16 V DC external (Nor340 adapter)6 x AA; external 5 to 15 V DC via I/O socket
Battery life (differs)8 h30 h
Weight (differs)380 g440 g
Dimensions (mm) (differs)215 x 74 x 29 (305 with microphone and preamplifier)280 x 80 x 42 (with preamplifier and microphone)
Operating temperature (°C)-10 to 50-10 to 50
Connectivity and software
Connectivity (differs)USB 2.0, analogue AC output (100 mV full scale)USB 2.0 Mini-B, 3.5 mm AC output, multipin I/O (AC and DC output)
Software (differs)NorXfer, NorReview, NorProtector, NorPower, NorConcertControl, Virtual InstrumentAnalyzerPlus

Change the selection

Choose up to 4 instruments of the same type. Instruments are grouped by brand; the type is shown after the model name.

Compare sound level meters | soundlevelmeter.review