Electroacoustics. Specifications for personal sound exposure meters

By soundlevelmeter.review editors, practising acousticians. Published . Neutral by policy: no brand pays for placement.

IEC 61252 specifies personal sound exposure meters, the instruments usually called noise dosimeters. It is the standard to look for on a dosimeter datasheet in the same way that IEC 61672-1 is for a sound level meter. The American counterpart is ANSI/ASA S1.25, which covers the same instrument but adds the exchange-rate, criterion and threshold settings used in United States regulation; most dosimeters sold internationally state conformance to both.

Editions

The first edition was published in 1993 and has been amended twice, in 2000 and 2017, giving the consolidated edition 1.2 that most current datasheets cite. A revised edition has been in preparation and is beginning to appear on the datasheets of newly announced instruments; where a manufacturer cites it, the statement is reproduced as given. The 2017 amendment aligned terminology with IEC 61672-1 and modernised test conditions without changing the essential tolerances.

What a personal sound exposure meter measures

The fundamental quantity is A-weighted sound exposure, the time integral of the squared A-weighted sound pressure over the measurement period, expressed in pascal-squared hours. Everything a dosimeter reports is derived from it: the daily noise exposure level LEX,8h used in ISO 9612 and European regulation, the time-weighted average and percentage dose used in American regulation, and the LAeq of the measurement period. An exposure of 1 Pa²h corresponds to an LEX,8h of about 85 dB(A). The standard requires the instrument to display sound exposure or a quantity derived from it, and to indicate overload.

Occupational regulations also limit the C-weighted peak sound pressure level, and nearly every dosimeter measures it. Peak measurement and the 5 dB exchange rate are features that regulations and ANSI S1.25 require rather than the core of IEC 61252, which is built on the equal-energy (3 dB) principle inherent in integrating sound exposure. The guide How to choose a noise dosimeter explains the settings.

Tolerances and tests

IEC 61252 sets tolerances on the A-weighting frequency response, on level linearity over the operating range, on the accuracy of time integration for steady and tone-burst signals, and on the effects of temperature, humidity, static pressure, vibration and electromagnetic fields. The tolerances correspond broadly to class 2 of IEC 61672-1, reflecting the fact that a shoulder-mounted microphone close to the body cannot deliver class 1 accuracy whatever the electronics do. The standard specifies the reference orientation and the acoustic test conditions for a microphone that is intended to be worn rather than pointed at a source, and it requires the manufacturer to publish the effect of the body and clothing where known.

IEC 61252 and IEC 61672-1 compared
AspectIEC 61252IEC 61672-1
InstrumentPersonal sound exposure meter, worn on the bodySound level meter, hand-held or mounted
Primary quantityA-weighted sound exposure (Pa²h)Time-weighted level, Leq, sound exposure level, peak
Frequency weightingsA (C for peak in practice)A, C, Z
Tolerance gradeOne grade, broadly equivalent to class 2Class 1 and class 2
Microphone positionOn the shoulder, near the earPointed at the source, away from the body
American counterpartANSI/ASA S1.25ANSI/ASA S1.4

Relationship to IEC 61672-1 and ISO 9612

A dosimeter is not a sound level meter under IEC 61672-1 unless the manufacturer also claims that standard, and several do, usually at class 2, so that the same instrument can be used for spot measurements. ISO 9612, the measurement standard for occupational exposure, accepts either a dosimeter to IEC 61252 or a sound level meter to IEC 61672-1 class 1 or 2 and sets out three strategies, task-based, job-based and full-day, for using them. It also requires an acoustic calibrator check before and after each measurement and an uncertainty estimate that includes instrument, sampling and microphone-position contributions.

Calibration

IEC 61252 has no separate periodic-test part like IEC 61672-3. Laboratories calibrate dosimeters to the manufacturer’s procedure or to the requirements of IEC 61252 itself, reporting the calibration level, the frequency response and the linearity with uncertainties. Field checks use an acoustic calibrator to IEC 60942 with an adaptor for the small microphone; badge-type instruments often dock in a reader that contains a calibrator. See Sound level meter calibration and verification.

Instruments stating conformance

The table lists every instrument in the database whose manufacturer-published standards list mentions IEC 61252, including sound level meters with an optional dosimetry function, quoting the edition where the manufacturer states it. Intrinsically safe variants and exchange-rate options are shown on each instrument page and in the dosimeter table.

Instruments table

26 instruments in the database citing IEC 61252, sorted by brand and model
InstrumentBrandTypeConformance as stated by the manufacturer
Bedrock Elite i10 BedrockHandheld sound level meterIEC 61252 (dosimeter)
dBadge CEL-35x DiscontinuedCasellaNoise dosimeterIEC 61252:2002; BS EN 61252:1997
dBadge2 CasellaNoise dosimeterIEC 61252 Ed. 1.1b:2002
doseBadge5 Personal Noise Dosimeter (CR:120A) Cirrus ResearchNoise dosimeterIEC 61252:1993 +AMD1:2000 (Ed 1.1, 2002-03)
Optimus+ Industrial Sound Level Meter Cirrus ResearchHandheld sound level meterIEC 61252:1993
Original doseBadge (CR:110A) Cirrus ResearchNoise dosimeterIEC 61252:1993
The doseBadge (CR:130A) AnnouncedCirrus ResearchNoise dosimeterIEC 61252:2025
Brüel & Kjær 4448 DiscontinuedHBK (Brüel & Kjær)Noise dosimeterIEC 61252:2002
SoundExpert 821ENV Larson DavisHandheld sound level meterIEC 61252:2017 (optional dosimetry)
Spartan 730 Larson DavisNoise dosimeterIEC 61252 Ed. 1.2
Spartan 730IS Larson DavisNoise dosimeterIEC 61252 Ed. 1.2
Spartan 821IH Larson DavisHandheld sound level meterIEC 61252:2017
NoiseBadge (Model 22) Pulsar InstrumentsNoise dosimeterIEC 61252:1993
NoiseBadge Intrinsically Safe (Model 22) Pulsar InstrumentsNoise dosimeterIEC 61252:1993
NoisePen Pulsar InstrumentsNoise dosimeterIEC 61252:1993
Nova Pulsar InstrumentsHandheld sound level meterIEC 61252:1993
Rion NB-14 RionNoise dosimeterIEC 61252:1993+AMD1:2000+AMD2:2017
Svantek SV 102A+ SvantekNoise dosimeterIEC 61252
Svantek SV 104 DiscontinuedSvantekNoise dosimeterIEC 61252 ed1.1 (2002)
Svantek SV 104A SvantekNoise dosimeterIEC 61252 Ed 1.2 (2017)
Svantek SV 104BIS SvantekNoise dosimeterIEC 61252 Ed 1.2 (R2017)
Svantek SV 104IS DiscontinuedSvantekNoise dosimeterIEC 61252 ed1.1 (2002)
Quest Edge 4+ (EG4P) TSI QuestNoise dosimeterIEC 61252 (personal sound exposure meters)
Quest Edge 7 (EG7) TSI QuestNoise dosimeterIEC 61252
Quest Edge 8 (EG8) TSI QuestNoise dosimeterIEC 61252:1993 + A1:2000 + A2:2017
TSI OmniWear Noise (OW-N-DVE) TSI QuestNoise dosimeterIEC 61252:1997 + A2:2017 (BS EN)