Skip to content
soundlevelmeter.review

CESVA SC250 vs B&K 2245

Every specification CESVA and HBK (Brüel & Kjær) publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationCESVA SC250CESVAB&K 2245HBK (Brüel & Kjær)
StatusCurrentCurrent
Verified on
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2013 class 1, EN 61672-1:2013 class 1, ANSI S1.4:2014 part 1 class 1, ANSI S1.43-1997 (R2007) class 1, IEC 61260-1:2014 class 1, EN 61260-1:2014 class 1, ANSI S1.11-2004 (R2009) class 1, OIML R 58:1998, OIML R 88:1998, OIML R 130:2001, DIN 45657:2014 (Taktmaximalpegel), IEC 61010-1:2010 A1:2016 / EN 61010-1:2010 A1:2019IEC 61672-1:2013 class 1, IEC 61260-1:2014 class 1 (with frequency analysis licence), ANSI/ASA S1.4-2014 class 1, ANSI/ASA S1.11-2014 class 1 (with frequency analysis licence)
Pattern approvals (differs)Spain (type examination 14/09/2021; additionals 24/02/2022, 18/04/2023, 04/06/2024), Portugal (30/03/2022), Czech Republic (22/06/2022), Switzerland (07/12/2022)PTB 1.1.2.386, certificate DE-20-M-PTB-0026
Measurement
Channels11
Measurement range (differs)Single range, 16.5 dB(A) to 140.1 dB(C) peak22.7 to 140.6 dB(A) single range
Linearity range (differs)—23.2 to 140.3 dB(A) at 1 kHz (IEC 61672); 19.8 to 141.5 dB(A) (IEC 60804)
Peak range (differs)—43.3 to 143.3 dB (C-weighted peak at 1 kHz)
Frequency range (differs)—6 Hz to 20 kHz
Frequency weightings (differs)—A, B, C, Z
Time weightings (differs)—F, S, I
Simultaneous weightings (differs)—No
Parameters (differs)F1, F2 and F3 user-selected functions, NC and NR curve evaluation, pass / not pass against an adjustable threshold on F1LXY, LXYmax, LXYmin, LXpeak, LXeq, LAIeq, LAImax, LXE, LAN statistics (five percentiles), LAFTeq, LavS4, LavS5
Analysis
Octave bands1/1 and 1/31/1 and 1/3
Octave band filter class (differs)1/1 octave 8 Hz to 16 kHz; 1/3 octave 6.3 Hz to 20 kHz (optional module)class 1 (IEC 61260-1:2014), one weighting
Reverberation time (differs)—No
Building acoustics (differs)—No
Audio recording (differs)—Yes
Logging and data
Shortest logging interval (differs)—1 s
Memory (differs)Built in; recordings and final results saved as CSV, with a log of date, sensitivity and firmware changes16 GB internal (about 13 GB formatted)
Hardware
Microphone (differs)—1/2" prepolarised free-field, Type 4966, 50 mV/Pa
Display (differs)3.5 inch capacitive colour touch screen240 x 320 transflective backlit colour LCD, 360-degree status light ring
Battery (differs)—Li-ion, 6700 mAh (24 Wh)
Battery life (differs)—13 h
Weight (differs)—370 g
Dimensions (mm) (differs)—260 x 68 x 37 including preamplifier and microphone
IP rating (differs)—IP55
Operating temperature (°C) (differs)—-25 to 70
Connectivity and software
Connectivity (differs)USB-C, Bluetooth, WiFi (CESVACloud, firmware updates, module activation)USB-C, Ethernet over USB, WiFi 2.4 GHz, Bluetooth Low Energy, GPS
Remote control (differs)—Noise Partner family mobile (iOS, Android) and PC apps; Open Interface BZ-7400
Software (differs)CESVA Lab, Result Merging Tools (SE003), SC250 Link (Android)Noise Partner (BZ-7300), Enviro Noise Partner (BZ-7301), Work Noise Partner (BZ-7302), Product Noise Partner (BZ-7303), Exhaust Noise Partner (BZ-7304), Open Interface (BZ-7400)

The instruments on their own