Spartan 821IH vs LA-4441A
Every specification Larson Davis and Ono Sokki publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.
| Specification | Spartan 821IHLarson Davis | LA-4441AOno Sokki |
|---|---|---|
| Status | Current | Current |
| Verified on | ||
| Standards and class | ||
| IEC 61672-1 class | Class 1 | Class 1 |
| Standards (differs) | IEC 61672-1:2013 class 1, ANSI S1.4-2014 class 1, IEC 61252:2017, ANSI S1.25-1991 | IEC 61672-1:2013 class 1, JIS C 1509-1:2017 class 1, JIS C 1516:2020 class 1 |
| Measurement | ||
| Channels | 1 | 1 |
| Measurement range (differs) | 17 to 140 dB(A) typical dynamic range | A: 27 to 135 dB; C: 30 to 135 dB; Z: 35 to 135 dB |
| Linearity range (differs) | — | 100 dB (wide range), 80 dB (normal range) |
| Frequency range (differs) | — | 10 Hz to 20 kHz |
| Frequency weightings | A, C, Z | A, C, Z |
| Time weightings | F, S, I | F, S, I |
| Parameters (differs) | Leq, Lmax, Lmin, Lpeak, Ln (6 values), Dose, Projected Dose, TWA, Projected TWA, LAIeq | Lp, Leq, LE, Lmax, Lmin, Lpeak, LN (Lhigh, L5, L10, L50, L90, L95, Llow, Lave) |
| Dosimetry (differs) | Yes | — |
| Exchange rates (differs) | 3 dB, 5 dB, custom 3 to 6 dB | — |
| Analysis | ||
| Octave bands (differs) | 1/1 and 1/3 | None |
| Octave band filter class (differs) | 6.3 Hz to 20 kHz (option X21-OB3 / -D models) | — |
| Audio recording (differs) | — | No |
| Logging and data | ||
| Shortest logging interval (differs) | 1 s (time history); measurement history 1 min to 6 h | 1 ms (Lp record memory: 1, 2, 5 or 10 ms) |
| Memory (differs) | 8 GB industrial microSD standard, up to 256 GB | Lp record 1000 addresses x 200 blocks; Leq 1000 addresses x 200 blocks; total values 100 addresses x 200 blocks; 9 condition memories |
| Hardware | ||
| Microphone (differs) | 1/2" prepolarised free-field 377B02 (821); 375A04 (721); preamplifier PRM821 or PRM721 | MI-1235 1/2" back-electret condenser, -29 dB nominal sensitivity, with MI-3111 preamplifier |
| Detachable preamplifier | Yes | Yes |
| Display (differs) | multi-colour back-lit LCD touch screen | 124 x 64 dot LCD with LED backlight |
| Battery (differs) | internal rechargeable; charging by USB-C (5 V, 3 A) or Qi wireless pad | 2 x AA (alkaline); optional PB-7090 AC adapter |
| Battery life (differs) | — | 8 h |
| Weight (differs) | 382.5 g | 360 g |
| Dimensions (mm) (differs) | 257.2 x 82.6 x 32.7 (with microphone and preamplifier); body 203.2 x 82.6 x 32.7 | 78 x 263 x 47 |
| Operating temperature (°C) (differs) | -20 to 60 | -10 to 50 |
| Connectivity and software | ||
| Connectivity (differs) | USB-C (USB 2.0), Bluetooth | USB 2.0 Full Speed (virtual COM port), RS-232C, AC output, DC output, external control input |
| Remote control (differs) | G4 LD Utility; LD Atlas app | — |
| Software (differs) | G4 LD Utility, LD Atlas | — |
The instruments on their own
- Spartan 821IH — full record, sources and owner reviews
- LA-4441A — full record, sources and owner reviews
Other comparisons
Want a different combination, or more than two at once? Build your own comparison from any instruments of the same type.