doseBadge5 Personal Noise Dosimeter (CR:120A) vs NoiseBadge (Model 22)
Every specification Cirrus Research and Pulsar Instruments publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.
| Specification | doseBadge5 Personal Noise Dosimeter (CR:120A)Cirrus Research | NoiseBadge (Model 22)Pulsar Instruments |
|---|---|---|
| Status | Current | Current |
| Verified on | ||
| Manufacturer list price | GBP 510 | |
| Standards and class | ||
| Standards (differs) | IEC 61252:1993 +AMD1:2000 (Ed 1.1, 2002-03), ANSI S1.25:1991 (R2017), IEC 61260-1:2014 class 2 and ANSI S1.11-2014 class 2 (octave bands) | IEC 61252:1993, ANSI S1.25:1991, IEC 60942:2001 class 2 (reader calibrator) |
| Recommended calibration interval (differs) | — | 12 months |
| Measurement | ||
| Channels (differs) | 6 | 2 |
| Measurement range (differs) | 60 to 140 dB(A) RMS; linear operating range 65 to 140 dB(A) | 70 to 130 dB(A) RMS |
| Linearity range (differs) | 65 to 140 dB(A) | — |
| Peak range (differs) | 80 to 143 dB(C) | 120 to 140 dB(C) |
| Frequency range (differs) | 63 Hz to 8 kHz (octave bands) | — |
| Frequency weightings (differs) | A, C, Z | A, C |
| Time weightings (differs) | F, S | S |
| Simultaneous weightings (differs) | Yes | — |
| Parameters (differs) | Leq/LAVG, LEP,d / LEX,8h / TWA, %Dose and estimated %Dose, Lpeak, sound exposure, ULT duration, SPL max and min, LCeq-LAeq (HML) | LAeq, LEP,d / LEX,8h, LAE, LCpeak, Lavg, TWA, Dose %, estimated Dose %, exposure Pa2h, 115 dB(A) exceedance |
| Dosimetry | Yes | Yes |
| Exchange rates | 3 dB, 4 dB, 5 dB | 3 dB, 4 dB, 5 dB |
| Analysis | ||
| Octave bands (differs) | 1/1 | None |
| Octave band filter class (differs) | class 2 (IEC 61260-1:2014) | — |
| Reverberation time (differs) | — | No |
| Building acoustics (differs) | — | No |
| Audio recording (differs) | — | No |
| Logging and data | ||
| Shortest logging interval (differs) | 1 s | 1 min |
| Memory (differs) | up to 80 h of time history (6 channels); up to 40 measurements; 24 h maximum single measurement | reader stores up to 93 eight-hour, 64 twelve-hour or 33 twenty-four-hour measurements |
| Hardware | ||
| Microphone (differs) | — | removable prepolarised electret, 1/2" class 2 |
| Detachable preamplifier | No | No |
| Display (differs) | none | none on badge (status LED); backlit LCD on reader |
| Battery (differs) | internal NiMH | internal NiMH, about 3 h charge; reader 2 x AA or USB |
| Battery life | 20 h | 20 h |
| External power (differs) | charging dock (3 h from empty) | — |
| Weight (differs) | 85 g | 45 g |
| Dimensions (mm) (differs) | 66 x 43 x 53 (excluding clips) | height 38, base diameter 47, microphone apex diameter 13 |
| Operating temperature (°C) | -10 to 50 | -10 to 50 |
| Connectivity and software | ||
| Connectivity (differs) | Bluetooth, USB via charging dock | infrared to reader, USB 2.0 (reader to PC) |
| Remote control (differs) | doseBadge5 remote control wand; dBLink / MyCirrus app (iOS, Android) | — |
| Software (differs) | NoiseTools, dBLink app | AnalyzerPlus |
The instruments on their own
- doseBadge5 Personal Noise Dosimeter (CR:120A) — full record, sources and owner reviews
- NoiseBadge (Model 22) — full record, sources and owner reviews
Other comparisons
- dBadge CEL-35x vs doseBadge5 Personal Noise Dosimeter (CR:120A)
- doseBadge5 Personal Noise Dosimeter (CR:120A) vs Spartan 730
- doseBadge5 Personal Noise Dosimeter (CR:120A) vs Svantek SV 102A+
- doseBadge5 Personal Noise Dosimeter (CR:120A) vs TES-660
- doseBadge5 Personal Noise Dosimeter (CR:120A) vs Quest Edge 4+ (EG4P)
- NoiseBadge (Model 22) vs Quest Edge 4+ (EG4P)
- doseBadge5 Personal Noise Dosimeter (CR:120A) vs Quest Edge 5 (EG5)
- doseBadge5 Personal Noise Dosimeter (CR:120A) vs Quest NoisePro DLX (NP-DLX)
Want a different combination, or more than two at once? Build your own comparison from any instruments of the same type.