doseBadge5 Personal Noise Dosimeter (CR:120A) vs TES-660
Every specification Cirrus Research and TES Electrical Electronic publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.
| Specification | doseBadge5 Personal Noise Dosimeter (CR:120A)Cirrus Research | TES-660TES Electrical Electronic |
|---|---|---|
| Status | Current | Current |
| Verified on | ||
| Standards and class | ||
| Standards (differs) | IEC 61252:1993 +AMD1:2000 (Ed 1.1, 2002-03), ANSI S1.25:1991 (R2017), IEC 61260-1:2014 class 2 and ANSI S1.11-2014 class 2 (octave bands) | IEC 61252:2017, ANSI S1.25:1991 |
| Measurement | ||
| Channels (differs) | 6 | 1 |
| Measurement range (differs) | 60 to 140 dB(A) RMS; linear operating range 65 to 140 dB(A) | 70 to 140 dB |
| Linearity range (differs) | 65 to 140 dB(A) | — |
| Peak range (differs) | 80 to 143 dB(C) | 103 to 143 dB |
| Frequency range (differs) | 63 Hz to 8 kHz (octave bands) | — |
| Frequency weightings | A, C, Z | A, C, Z |
| Time weightings (differs) | F, S | F, S, I |
| Simultaneous weightings (differs) | Yes | — |
| Parameters (differs) | Leq/LAVG, LEP,d / LEX,8h / TWA, %Dose and estimated %Dose, Lpeak, sound exposure, ULT duration, SPL max and min, LCeq-LAeq (HML) | Lavg, LZpk, Leq, LCpk |
| Dosimetry | Yes | Yes |
| Exchange rates (differs) | 3 dB, 4 dB, 5 dB | 3 dB, 4 dB, 5 dB, 6 dB |
| Intrinsically safe marking (differs) | — | TES-660A only: ATEX, FM/CSA; EN 60079-0:2006, EN 60079-11:2007, EN 60079-26:2007 |
| Analysis | ||
| Octave bands (differs) | 1/1 | — |
| Octave band filter class (differs) | class 2 (IEC 61260-1:2014) | — |
| Logging and data | ||
| Shortest logging interval (differs) | 1 s | — |
| Memory (differs) | up to 80 h of time history (6 channels); up to 40 measurements; 24 h maximum single measurement | 30 measurements |
| Hardware | ||
| Microphone (differs) | — | 1/2" electret, field-replaceable, threaded |
| Detachable preamplifier (differs) | No | — |
| Display (differs) | none | 128 x 64 graphic |
| Battery (differs) | internal NiMH | Rechargeable Li-polymer 3.7 V 470 mAh |
| Battery life (differs) | 20 h | — |
| External power (differs) | charging dock (3 h from empty) | — |
| Weight (differs) | 85 g | 77 g |
| Dimensions (mm) (differs) | 66 x 43 x 53 (excluding clips) | 84 x 49 x 55 (with windscreen) |
| Operating temperature (°C) (differs) | -10 to 50 | 0 to 50 |
| Connectivity and software | ||
| Connectivity (differs) | Bluetooth, USB via charging dock | USB via DS-01 charging unit |
| Remote control (differs) | doseBadge5 remote control wand; dBLink / MyCirrus app (iOS, Android) | — |
| Software (differs) | NoiseTools, dBLink app | — |
The instruments on their own
- doseBadge5 Personal Noise Dosimeter (CR:120A) — full record, sources and owner reviews
- TES-660 — full record, sources and owner reviews
Other comparisons
- dBadge CEL-35x vs doseBadge5 Personal Noise Dosimeter (CR:120A)
- doseBadge5 Personal Noise Dosimeter (CR:120A) vs Spartan 730
- Spartan 730 vs TES-660
- doseBadge5 Personal Noise Dosimeter (CR:120A) vs NoiseBadge (Model 22)
- doseBadge5 Personal Noise Dosimeter (CR:120A) vs Svantek SV 102A+