Every specification Larson Davis and Svantek publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationSoundAdvisor 831CLarson DavisSvantek SV 971ASvantek
StatusCurrentCurrent
Verified on
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2013 class 1, IEC 60651 type 1, IEC 60804 type 1, ANSI S1.4-2014 class 1, ANSI S1.43-1997 type 1, DIN 45657, IEC 61260:2014 class 1 (option 831C-OB3), ANSI S1.11-2014 class 1 (option 831C-OB3)IEC 61672-1:2013 class 1, IEC 61260-1:2014 class 1, ANSI/ASA S1.4, IEC 60268-16 (STIPA option)
Pattern approvals (differs)LNE, PTB, METAS (with PRM2103), METAS (with PRM831)PTB (Germany), BEV (Austria), GUM Central Office of Measures (Poland)
Measurement
Channels11
Measurement range (differs)single range; dynamic range > 120 dB; peak overload 143 dB with 50 mV/Pa microphone at 0 dB gain17 dB(A) Leq to 140 dB(A) peak (total range)
Linearity range (differs)—24 dB(A) Leq to 140 dB(A) peak
Peak range (differs)up to 143 dB peak (50 mV/Pa microphone, 0 dB gain)up to 140 dB(A) peak
Frequency range (differs)—5 Hz to 20 kHz
Frequency weightings (differs)A, C, ZA, B, C, Z
Time weightingsF, S, IF, S, I
Simultaneous weightings (differs)—Yes
Parameters (differs)Leq, Lmax, Lmin, Lpeak, Ln (6 selectable), Ldn, Lden, LCeq-LAeqLeq, SPL, Lmax, Lmin, Lpeak, SEL, Ln (L1 to L99), Dose, TWA, three parallel profiles
Dosimetry (differs)—Yes
Exchange rates (differs)—2 dB, 3 dB, 4 dB, 5 dB, 6 dB
Analysis
Octave bands1/1 and 1/31/1 and 1/3
Octave band filter class (differs)class 1 (IEC 61260:2014), 6.3 Hz to 20 kHz, option 831C-OB3class 1 (IEC 61260-1:2014), licensed option
FFT (differs)option 831C-FFT (FFT analysis and tone assessment); line count not published on the pages read—
Reverberation timeYesYes
Building acoustics (differs)—Yes
Speech intelligibility (differs)—STIPA (option)
Audio recordingYesYes
Logging and data
Shortest logging interval (differs)2.5 ms (option 831C-LOG); minimum integration time 0.1 s100 ms (2 ms in RT60 mode)
Memory (differs)2 GB internal, expandable via USB memory (831-MEMUSB)microSD up to 32 GB
Hardware
Microphone (differs)1/2" prepolarised free-field 377B02 (50 mV/Pa) or random-incidence 377C20; preamplifier PRM8311/2" ACO 7152 (SV 7152) with SV 18A preamplifier
Detachable preamplifierYesYes
Outdoor microphone kit (differs)PRM2103 outdoor preamplifier with EPS2116 environmental shroud—
Weather sensors (differs)Yes—
Display (differs)240 x 320 colour graphic LCD with touch screenColour display with 8-key keypad (size not stated)
Battery (differs)4 x AA (NiMH 2500 mAh supplied; alkaline or 1.5 V lithium also usable); external 10 to 25 V DC or 5 V via USB4 x AAA
Battery life (differs)8 h16 h
Weight (differs)532 g225 g
Dimensions (mm) (differs)290 x 71 x 41 (with microphone and preamplifier); body 224 long232.5 x 56 x 20
Operating temperature (°C) (differs)-30 to 60-10 to 50
Connectivity and software
Connectivity (differs)USB Mini-B, Ethernet (831-INT-ET or DVX012 adapter), WiFi (DVX014 adapter), cellular gateway (option 831C-SW, Sierra Wireless RV50X), GPS (GPS002 receiver), AC output, I/O connectorUSB-C, Bluetooth, WiFi/LAN via SvanLINK
Remote control (differs)built-in web interface from PC or mobile device; G4 LD Utility; LD Atlas appAssistant Pro app; Building Acoustics app; SvanLINK telemetry
Software (differs)G4 LD Utility, LD Atlas, DNAAssistant Pro app, Building Acoustics Assistant app, SvanPC++
Latest firmware (differs)5.4 (version supported by reference manual I831C.01(S), 2025)—

The instruments on their own