Skip to content
soundlevelmeter.review

Norsonic Nor150 vs Nova

Every specification Norsonic and Pulsar Instruments publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationNorsonic Nor150NorsonicNovaPulsar Instruments
StatusCurrentCurrent
Verified on
Manufacturer list priceGBP 1,100
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2013 class 1, IEC 61260-1 class 1, IEC 60651, IEC 60804, DIN 45657, ANSI S1.4, ANSI S1.11, ANSI S1.43, IEC 60268-16 (STIPA, option 17), ISO 16283, ISO 717IEC 61672-1:2013 class 1 or class 2, group X, IEC 61672-2:2013, IEC 60651:2001 type 1 I or type 2 I, IEC 60804:2000 type 1 or type 2, IEC 61252:1993, ANSI S1.4-1983 (R2006), ANSI S1.43-1997 (R2007), ANSI S1.25:1991, IEC 61260 and ANSI S1.11-2004 (1/1 octave filters)
Pattern approvals (differs)PTB (Germany), rev. 4, LNE 32377-4 (France), METAS (Switzerland), BEV (Austria), CIM (Czech Republic)—
Recommended calibration interval (differs)—12 months
Measurement
Channels (differs)21
Measurement range (differs)17 to 137 dB(A) RMS, single 120 dB range, plus 10 dB extended range20 to 140 dB RMS, single range
Peak range (differs)45 to 140 dB (C-weighted peak), plus 10 dB extensionup to 143 dB
Frequency range (differs)0.4 Hz to 20 kHz (option 3); to 40 kHz with ultrasound option 14—
Frequency weightings (differs)A, C, Z, B, G, UA, C, Z
Time weightingsF, S, IF, S, I
Simultaneous weightingsYesYes
Parameters (differs)Leq, Lmax, Lmin, LE, Lpeak, LFTm5, Ln percentiles, LAeq-LCeq, dual-channel synchronous measurementLxy, Lxy max/min, Lxeq, LCpeak, LZpeak, LApeak, LCeq-LAeq, LxE, LAIeq, Ln (14 values), TWA, Dose %, estimated Dose %
Dosimetry (differs)—Yes
Exchange rates (differs)—3 dB, 4 dB, 5 dB
Analysis
Octave bands (differs)1/1 and 1/31/1
Octave band filter class (differs)class 1 (IEC 61260), option 3IEC 61260 and ANSI S1.11-2004, 31.5 Hz to 16 kHz (Models 45/46)
FFT (differs)option 13; 8192 lines, 1.46 Hz resolution (16384 lines with ultrasound)—
Reverberation time (differs)YesNo
Building acoustics (differs)YesNo
Speech intelligibility (differs)STIPA (option 17)—
Audio recordingYesYes
Logging and data
Shortest logging interval (differs)5 ms10 ms (time history); periodic 1 min to 1 h and Lden
Memory (differs)SD card (tested to 512 GB)4 GB removable memory card
Hardware
Microphone (differs)1/2" Nor1225 microphone with Nor1209 preamplifier; Nor1290 intensity probe optional1/2" prepolarised PM1 (class 1) or PM2 (class 2); PA40 removable preamplifier
Detachable preamplifierYesYes
Outdoor microphone kit (differs)Nor1217 or Nor1216 outdoor microphonesWK3-N outdoor kit with PM172 outdoor microphone
Display (differs)4.3" colour touchscreen plus push keysOLED with ambient light sensor (product description says TFT)
Battery (differs)Replaceable Li-ion pack (NCA103450 cells); 9 to 15 V DC input6 x AA; external 5 to 15 V DC via I/O socket
Battery life (differs)8 h30 h
Weight (differs)700 g440 g
Dimensions (mm) (differs)240 x 82 x 39 excluding microphone280 x 80 x 42 (with preamplifier and microphone)
Operating temperature (°C) (differs)—-10 to 50
Connectivity and software
Connectivity (differs)LAN, USB host, digital I/O, GPS (option 1), built-in camera (option 1)USB 2.0 Mini-B, 3.5 mm AC output, multipin I/O (AC and DC output)
Remote control (differs)NorCloud (option 12), web server (option 11), Remote API (option 15)—
Software (differs)NorCloud, NorConnect, NorVirtual, Nor850 Suite, Nor850/SI Sound IntensityAnalyzerPlus
Latest firmware (differs)6.0.3025 (25 June 2026); release note v6.0.2920 (22 May 2026)—

The instruments on their own