Skip to content
soundlevelmeter.review

CESVA SC420 vs Nova

Every specification CESVA and Pulsar Instruments publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationCESVA SC420CESVANovaPulsar Instruments
StatusCurrentCurrent
Verified on
Manufacturer list priceGBP 1,100
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2013 class 1, EN 61672-1:2013 class 1, ANSI S1.4:2014 part 1, ANSI S1.43-1997 (R2007), IEC 61260-1:2014 class 1, EN 61260-1:2014 class 1, ANSI S1.11-2004 (R2009), OIML R 58, OIML R 88, OIML R 130, DIN 45657:2014 (Taktmaximalpegel), IEC 61010-1 / EN 61010-1 (pollution degree II)IEC 61672-1:2013 class 1 or class 2, group X, IEC 61672-2:2013, IEC 60651:2001 type 1 I or type 2 I, IEC 60804:2000 type 1 or type 2, IEC 61252:1993, ANSI S1.4-1983 (R2006), ANSI S1.43-1997 (R2007), ANSI S1.25:1991, IEC 61260 and ANSI S1.11-2004 (1/1 octave filters)
Pattern approvals (differs)Spain (type examination 11/09/2013; latest 11/07/2022), Portugal (17/02/2014, renewed 05/03/2024)—
Recommended calibration interval (differs)—12 months
Measurement
Channels11
Measurement range (differs)—20 to 140 dB RMS, single range
Peak range (differs)—up to 143 dB
Frequency weightings (differs)—A, C, Z
Time weightings (differs)—F, S, I
Simultaneous weightings (differs)—Yes
Parameters (differs)—Lxy, Lxy max/min, Lxeq, LCpeak, LZpeak, LApeak, LCeq-LAeq, LxE, LAIeq, Ln (14 values), TWA, Dose %, estimated Dose %
Dosimetry (differs)—Yes
Exchange rates (differs)—3 dB, 4 dB, 5 dB
Analysis
Octave bands (differs)1/1 and 1/31/1
Octave band filter class (differs)1/1 octave 16 Hz to 16 kHz (OF420); 1/3 octave 10 Hz to 20 kHz (TF420) (optional module)IEC 61260 and ANSI S1.11-2004, 31.5 Hz to 16 kHz (Models 45/46)
FFT (differs)10,000 lines, 0 to 20 kHz, 2 Hz per line (FF420 module)—
Reverberation time (differs)—No
Building acoustics (differs)—No
Audio recordingYesYes
Logging and data
Shortest logging interval (differs)—10 ms (time history); periodic 1 min to 1 h and Lden
Memory (differs)Removable MicroSD; cards up to 32 GB tested by the manufacturer4 GB removable memory card
Hardware
Microphone (differs)—1/2" prepolarised PM1 (class 1) or PM2 (class 2); PA40 removable preamplifier
Detachable preamplifier (differs)—Yes
Outdoor microphone kit (differs)—WK3-N outdoor kit with PM172 outdoor microphone
Display (differs)—OLED with ambient light sensor (product description says TFT)
Battery (differs)—6 x AA; external 5 to 15 V DC via I/O socket
Battery life (differs)—30 h
Weight (differs)—440 g
Dimensions (mm) (differs)—280 x 80 x 42 (with preamplifier and microphone)
Operating temperature (°C) (differs)—-10 to 50
Connectivity and software
Connectivity (differs)USB, Bluetooth class 1, external input for automatic start/stopUSB 2.0 Mini-B, 3.5 mm AC output, multipin I/O (AC and DC output)
Software (differs)CESVA Lab, Result Merging Tools (SE003), CESVA SC420 Link (Android)AnalyzerPlus

The instruments on their own