SpecificationNorsonic Nor1545NorsonicNoiseLog_mobilSinus Messtechnik
StatusCurrentCurrent
Verified on
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2013 class 1 (Nor145), IEC 61260-1 class 1 (Nor145)IEC 61672 class 1, IEC 60651, IEC 60804, IEC 61260 class 0 (filters), MIL-STD 810F (shock)
Pattern approvals (differs)Nor145 approvals: PTB, LNE 37403-1, METAS, BEV, CIM, IPQ, GUMPTB 21.21 / 13.05, METAS CH-A3-15103-00
Measurement
Channels (differs)18
Measurement range (differs)17 to 137 dB(A) RMS (Nor145)dynamic range 120 dB
Peak range (differs)45 to 140 dB (C-weighted peak, Nor145)—
Frequency range (differs)6.3 Hz to 20 kHz standard (Nor145); 0.4 Hz to 40 kHz with optionsDC to 40 kHz at 8 channels (80 kHz optional)
Frequency weightingsA, C, ZA, C, Z
Time weightings (differs)F, S, I—
Simultaneous weightings (differs)Yes—
Parameters (differs)Leq, Lmax, Lmin, LE, Lpeak, Ln percentiles, event markers, noise direction (Nor1297)Lp, Lpeak, Leq (A, C, Z)
Analysis
Octave bands (differs)1/1 and 1/31/3
Octave band filter class (differs)class 1 (IEC 61260)class 0 (IEC 61260)
FFT (differs)Nor145 FFT optionFFT analysis (line count not stated on page)
Audio recordingYesYes
Logging and data
Shortest logging interval (differs)5 ms (Nor145)—
Memory (differs)SD card in Nor145500 GB HDD or SSD; SD/SDXC
Hardware
Microphone (differs)Nor1216 heated outdoor microphone with Nor1227 cartridge, or Nor1217 with Nor1239 cartridgeoutdoor microphones such as 41AN, WME952, MM255 with WS1, 46AE with RA0153
Detachable preamplifierYesYes
Outdoor microphone kit (differs)Nor1216 (permanent, heated, PTB approved) or Nor1217 (temporary)telescoping MIL masts; optional pneumatic mast
Weather sensorsYesYes
Display (differs)4.3" touchscreen of the Nor14510.4" TFT 1024 x 768 touch screen
Battery (differs)Internal backup battery; Tracer Nor344 Li-polymer battery option (Nor1545/B)internal battery, up to 15 to 16 h UPS; trailer with batteries and charger
Battery life (differs)7 h—
External power (differs)88 to 264 V AC; 9 to 28 V DC; solar panel; typical consumption 3.5 W100 to 240 V AC or 10 to 36 V DC; trailer 230 V AC / 24 V DC on-board supply
Autonomy on internal power (differs)3 days7 days
Weight (differs)3300 g6000 g
Dimensions (mm) (differs)200 x 132 x 400500 x 400 x 220 (case; Soundbook spec gives 600 x 400 x 220)
IP rating (differs)IP66 (IP67 with extra sealant in the cable gland)IP65 (case); IP54 (PC and trailer)
Operating temperature (°C) (differs)-20 to 50 (battery charging 0 to 40)-10 to 50
Connectivity and software
Connectivity (differs)LAN, WiFi, 3G/UMTS, 4G/LTE, GPS, digital I/O (Nor1545/B)LAN, WLAN, GPRS, Bluetooth, modem RJ-11, optional 3G/4G/LTE router, alarm relay outputs
Remote control (differs)NorCloudfull remote control over network; MUNISENSE web platform; e-mail and relay alarms
Software (differs)NorCloud, NorReviewSAMURAI, Auditor, MUNISENSE
Latest firmware (differs)Nor145 6.0.3025 (25 June 2026)—

The instruments on their own