Skip to content
soundlevelmeter.review

CESVA SC420 vs Optimus+ Environmental Sound Level Meter

Every specification CESVA and Cirrus Research publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationCESVA SC420CESVAOptimus+ Environmental Sound Level MeterCirrus Research
StatusCurrentCurrent
Verified on
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2013 class 1, EN 61672-1:2013 class 1, ANSI S1.4:2014 part 1, ANSI S1.43-1997 (R2007), IEC 61260-1:2014 class 1, EN 61260-1:2014 class 1, ANSI S1.11-2004 (R2009), OIML R 58, OIML R 88, OIML R 130, DIN 45657:2014 (Taktmaximalpegel), IEC 61010-1 / EN 61010-1 (pollution degree II)IEC 61672-1:2013 class 1 or class 2, IEC 61260 and ANSI S1.11-2004 (1:1 and 1:3 octave filters), IEC 60651, IEC 60804, ANSI S1.4, ANSI S1.43, DIN 45657
Pattern approvals (differs)Spain (type examination 11/09/2013; latest 11/07/2022), Portugal (17/02/2014, renewed 05/03/2024)PTB (class 1), LNE (class 1)
Recommended calibration interval (differs)—12 months
Measurement
Channels11
Measurement range (differs)—20 to 140 dB, single range
Frequency range (differs)—6.3 Hz to 20 kHz (1:3 octave bands measured; 12.5 Hz to 20 kHz displayed)
Frequency weightings (differs)—A, C, Z
Time weightings (differs)—F, S, I
Simultaneous weightings (differs)—Yes
Parameters (differs)—LXeq, LXYmax, LXYmin, LCpeak, LZpeak, LXE, Ln statistics, Lden, Leq LF (20 to 200 Hz), tonal noise detection (C models)
Analysis
Octave bands1/1 and 1/31/1 and 1/3
Octave band filter class (differs)1/1 octave 16 Hz to 16 kHz (OF420); 1/3 octave 10 Hz to 20 kHz (TF420) (optional module)—
FFT (differs)10,000 lines, 0 to 20 kHz, 2 Hz per line (FF420 module)—
Audio recordingYesYes
Logging and data
Shortest logging interval (differs)—10 ms
Memory (differs)Removable MicroSD; cards up to 32 GB tested by the manufacturer8 GB standard; 32 GB factory option
Hardware
Microphone (differs)—class 1: MK:224 1/2" pre-polarised; class 2: MK:216 pre-polarised
Detachable preamplifier (differs)—Yes
Outdoor microphone kit (differs)—Outdoor Environmental Noise Kit (CK:67x/68x) with MK:172 outdoor microphone
Display (differs)—high-resolution OLED with ambient light sensor and illuminated keypad
Battery (differs)—4 x AA alkaline or lithium
Battery life (differs)—12 h
External power (differs)—5 to 15 V DC via ZL:171 cable
Weight (differs)—300 g
Dimensions (mm) (differs)—283 x 65 x 30
Operating temperature (°C) (differs)—-10 to 50
Connectivity and software
Connectivity (differs)USB, Bluetooth class 1, external input for automatic start/stopUSB type B, Bluetooth Low Energy, AC/DC output, GPS (GPS model)
Remote control (differs)—MyCirrus mobile apps via Bluetooth; remote access via Outdoor Kit modem
Software (differs)CESVA Lab, Result Merging Tools (SE003), CESVA SC420 Link (Android)NoiseTools, MyCirrus app

The instruments on their own

Want a different combination, or more than two at once? Build your own comparison from any instruments of the same type.