Skip to content
soundlevelmeter.review

B&K 2250 vs Tango Plus

Every specification HBK (Brüel & Kjær) and Sinus Messtechnik publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationB&K 2250HBK (Brüel & Kjær)Tango PlusSinus Messtechnik
StatusDiscontinuedCurrent
Verified on
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2013 class 1, IEC 60651 type 1, IEC 60804 type 1, DIN 45657, ANSI S1.4 type 1, ANSI/ASA S1.4-2014 class 1, ANSI S1.43 type 1, IEC 61260-1:2014 class 1 (BZ-7223), ANSI/ASA S1.11-2014 class 1 (BZ-7223), ISO 7196:1995 (infrasound), ISO 8041:2005 (vibration)IEC 61672 class 1, IEC 60651, IEC 60804, IEC 61260 class 1 (1/1 and 1/3 octave filters)
Pattern approvals (differs)—PTB DE-20-M-PTB-0033
Measurement
Channels11
Measurement range (differs)about 16.6 to 140 dB(A) dynamic range with Type 4189; primary display range 23.5 to 122.3 dB(A)25 to 140 dB (LAeq)
Linearity range (differs)24.8 to 139.7 dB(A) at 1 kHz (IEC 61672); 21.4 to 140.8 dB(A) (IEC 60804)—
Peak range (differs)42.3 to 142.7 dB (C-weighted peak at 1 kHz)60 to 143 dB (LCpeak)
Frequency range (differs)0.5 Hz to 20 kHz broadband; 1/3-octave 6.3 Hz to 20 kHz (0.8 Hz with low-frequency option)10 Hz to 20 kHz (spectra)
Frequency weightings (differs)A, B, C, Z, GA, C, Z
Time weightingsF, S, IF, S, I
Parameters (differs)LXeq, LXE, Lmax, Lmin, LN statistics, Ldn, Lden, Lday, Levening, Lnight, NR, RC, NC, NCB, SIL, loudness, tonal assessment (Kt, dLta, Lpt)LAF, LAS, LCF, LAFmax, LASmax, LCFmax, LAFmin, LASmin, LCFmin, LAeq, LCeq, LCpeak, LAE,1s, LAE,8h, LAFT (LAtm5), LAFTeq, LAFn (3 percentiles), LCeq-LAeq, LAFTeq-LAeq, LAImax, LCpeak exceedance counts (3 levels)
Analysis
Octave bands1/1 and 1/31/1 and 1/3
Octave band filter class (differs)class 1 (IEC 61260-1:2014) with BZ-7223class 1 (IEC 61260), 10 Hz to 20 kHz
FFT (differs)BZ-7230 FFT option (line count not stated in BP 2029)—
Reverberation time (differs)Yes—
Building acoustics (differs)Yes—
Audio recording (differs)Yes—
Logging and data
Shortest logging interval (differs)100 ms fast logging (10 ms for LAF); 1 s standard62.5 ms
Memory (differs)512 MB internal flash; two SD/SDHC slots (up to 32 GB); USB stick32 MB
Hardware
Microphone (differs)1/2" prepolarised free-field, Type 4189, 50 mV/Pa (Types 4190 and 4966 alternatives)1/2" prepolarised MK255 class 1, 50 mV/Pa, removable, LEMO 7-pin
Detachable preamplifierYesYes
Display (differs)240 x 320 transflective backlit colour touchscreencolour TFT 320 x 240
Battery (differs)Li-ion pack QB-0061, 3.7 V, 5200 mAh2 x AA (1.2 to 1.5 V); USB powered from PC; optional outdoor kit with Li battery
Battery life (differs)11 h40 h
Weight (differs)650 g290 g
Dimensions (mm) (differs)300 x 93 x 50 including preamplifier and microphone190 x 70 x 32
Operating temperature (°C)-10 to 50-10 to 50
Connectivity and software
Connectivity (differs)USB 2.0 (OTG Micro-AB and Standard-A), LAN (RJ45, 100 Mbit/s), WiFi via UL-1050/UL-1019 adapter, USB modem (GPRS/EDGE/HSPA)USB 2.0, printer port, RS232 (optional adapter), 3.5 mm audio output
Remote control (differs)Measurement Partner Field App, Measurement Partner Cloud—
Software (differs)Measurement Partner Suite (BZ-5503), Sound Level Meter BZ-7222, Frequency Analysis BZ-7223, Logging BZ-7224, Enhanced Logging BZ-7225, Signal Recording BZ-7226, Reverberation Time BZ-7227, Building Acoustics BZ-7228, FFT BZ-7230, Tone Assessment BZ-7231, Noise Monitoring BZ-7232Tango-Utilities, AUDITOR_light, AUDITOR

The instruments on their own

Want a different combination, or more than two at once? Build your own comparison from any instruments of the same type.