Skip to content
soundlevelmeter.review

SoundTrack LxT vs Tango Plus

Every specification Larson Davis and Sinus Messtechnik publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationSoundTrack LxTLarson DavisTango PlusSinus Messtechnik
StatusDiscontinuedCurrent
Verified on
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2013 class 1 (LxT1) or class 2 (LxT2), ANSI S1.4-2014 class 1 or class 2, ANSI S1.43-1997 (R2002), IEC 61260:2001 class 1 / ANSI S1.11-2004 class 1 (octave filters, option), IEC 60651:2001, IEC 60804:2000, ISO 9612:2009 measurement strategies supportedIEC 61672 class 1, IEC 60651, IEC 60804, IEC 61260 class 1 (1/1 and 1/3 octave filters)
Pattern approvals (differs)—PTB DE-20-M-PTB-0033
Measurement
Channels11
Measurement range (differs)17 to 118 dB(A), 19 to 118 dB(C), 24 to 118 dB(Z) typical dynamic range (single range; LxT1 with 377B02)25 to 140 dB (LAeq)
Peak range (differs)—60 to 143 dB (LCpeak)
Frequency range (differs)1/1 octave 8 Hz to 16 kHz; 1/3 octave 6.3 Hz to 20 kHz (filter options)10 Hz to 20 kHz (spectra)
Frequency weightingsA, C, ZA, C, Z
Time weightingsF, S, IF, S, I
Parameters (differs)SPL, Leq, Lmax, Lmin, SEL, Lpeak, Lpeak(max), Ln (6 values L0.01 to L99.99), Lavg, TWA, Dose, ProjDose, Lep,d, E, E8, E40, SEALAF, LAS, LCF, LAFmax, LASmax, LCFmax, LAFmin, LASmin, LCFmin, LAeq, LCeq, LCpeak, LAE,1s, LAE,8h, LAFT (LAtm5), LAFTeq, LAFn (3 percentiles), LCeq-LAeq, LAFTeq-LAeq, LAImax, LCpeak exceedance counts (3 levels)
Dosimetry (differs)Yes—
Analysis
Octave bands1/1 and 1/31/1 and 1/3
Octave band filter class (differs)options LXT-OB1 (1/1) and LXT-OB3 (1/3 and 1/1)class 1 (IEC 61260), 10 Hz to 20 kHz
Logging and data
Shortest logging interval (differs)100 ms (option LXT-HSLOG); 1 s with LXT-LOG62.5 ms
Memory (differs)2 GB32 MB
Hardware
Microphone (differs)1/2" 377B02 free-field 50 mV/Pa (LxT1) or 375A04 (LxT2); optional 377C20 random-incidence; preamplifier PRMLXT1 or PRMLXT2C; drives up to 61 m extension cable1/2" prepolarised MK255 class 1, 50 mV/Pa, removable, LEMO 7-pin
Detachable preamplifierYesYes
Display (differs)160 x 240 pixel backlit LCDcolour TFT 320 x 240
Battery (differs)4 x AA (alkaline, NiMH or 1.5 V lithium); PSA029 mains supply; BAT015 external D-cell holder2 x AA (1.2 to 1.5 V); USB powered from PC; optional outdoor kit with Li battery
Battery life (differs)16 h40 h
External power (differs)USB/power port; optional BAT015 external D-cell pack (over 13 days)—
Weight (differs)513 g290 g
Dimensions (differs)290 x 71 x 41 overall (body 224 mm long)190 x 70 x 32
Operating temperature (°C) (differs)-10 to 50 (sensitivity within +/-0.5 dB); storage -30 to 60-10 to 50
Connectivity and software
Connectivity (differs)USB 2.0, auxiliary USB for storage, AC/DC output and headset jackUSB 2.0, printer port, RS232 (optional adapter), 3.5 mm audio output
Software (differs)G4 LD Utility (SWW-SLM-UTILG4), Blaze (SWW-BLAZE-LXT), DNATango-Utilities, AUDITOR_light, AUDITOR
Latest firmware (differs)2.404 (reference manual revision R)—

The instruments on their own

Want a different combination, or more than two at once? Build your own comparison from any instruments of the same type.