Skip to content
soundlevelmeter.review

LA-3570 vs Nova

Every specification Ono Sokki and Pulsar Instruments publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationLA-3570Ono SokkiNovaPulsar Instruments
StatusDiscontinuedCurrent
Verified on
Manufacturer list priceGBP 1,100
Standards and class
IEC 61672-1 classClass 1Class 1
Standards (differs)IEC 61672-1:2002 class 1, JIS C 1509-1:2005 class 1, IEC 61260:1995 class 1 (octave options LA-0351/LA-0352)IEC 61672-1:2013 class 1 or class 2, group X, IEC 61672-2:2013, IEC 60651:2001 type 1 I or type 2 I, IEC 60804:2000 type 1 or type 2, IEC 61252:1993, ANSI S1.4-1983 (R2006), ANSI S1.43-1997 (R2007), ANSI S1.25:1991, IEC 61260 and ANSI S1.11-2004 (1/1 octave filters)
Recommended calibration interval (differs)—12 months
Measurement
Channels11
Measurement range (differs)A: 22 to 130 dB; C: 28 to 130 dB; Z: 36 to 130 dB20 to 140 dB RMS, single range
Peak range (differs)—up to 143 dB
Frequency range (differs)10 Hz to 15 kHz—
Frequency weightingsA, C, ZA, C, Z
Time weightingsF, S, IF, S, I
Simultaneous weightings (differs)—Yes
Parameters (differs)Lp, Leq, LE, Lpeak, Lmax, Lmin, LN (L5, L10, L50, L90, L95 and others), Loudness DIN 45631 (option LA-0358)Lxy, Lxy max/min, Lxeq, LCpeak, LZpeak, LApeak, LCeq-LAeq, LxE, LAIeq, Ln (14 values), TWA, Dose %, estimated Dose %
Dosimetry (differs)—Yes
Exchange rates (differs)—3 dB, 4 dB, 5 dB
Analysis
Octave bands (differs)1/1 and 1/31/1
Octave band filter class (differs)class 1 (IEC 61260:1995), options LA-0351 (1/1) and LA-0352 (1/3)IEC 61260 and ANSI S1.11-2004, 31.5 Hz to 16 kHz (Models 45/46)
FFT (differs)400 lines (option LA-0353)—
Reverberation time (differs)—No
Building acoustics (differs)—No
Audio recordingYesYes
Logging and data
Shortest logging interval (differs)10 ms (option LA-0356)10 ms (time history); periodic 1 min to 1 h and Lden
Memory (differs)SD/SDHC card up to 32 GB (1 GB card supplied)4 GB removable memory card
Hardware
Microphone (differs)MI-1211 1/2" externally polarised (bias type), -20 dB nominal sensitivity, with MI-3310 preamplifier1/2" prepolarised PM1 (class 1) or PM2 (class 2); PA40 removable preamplifier
Detachable preamplifierYesYes
Outdoor microphone kit (differs)—WK3-N outdoor kit with PM172 outdoor microphone
Display (differs)3.5" LCD with white backlightOLED with ambient light sensor (product description says TFT)
Battery (differs)4 x AA (alkaline or rechargeable); AC adapter PB-70906 x AA; external 5 to 15 V DC via I/O socket
Battery life (differs)8 h30 h
Weight (differs)680 g440 g
Dimensions (mm) (differs)379 x 106 x 49.3280 x 80 x 42 (with preamplifier and microphone)
Operating temperature (°C)-10 to 50-10 to 50
Connectivity and software
Connectivity (differs)USB (mass storage, mini-B), RS-232C, SD card, AC output, AC/DC output, headphone output, external control input, comparator outputUSB 2.0 Mini-B, 3.5 mm AC output, multipin I/O (AC and DC output)
Software (differs)OS-2000 series, DS-3000 seriesAnalyzerPlus

The instruments on their own

Want a different combination, or more than two at once? Build your own comparison from any instruments of the same type.