Skip to content
soundlevelmeter.review

HD2010UC vs Quest SoundPro SE/DL

Every specification Delta OHM (Senseca) and TSI Quest publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.

Side-by-side specifications. Rows where the instruments differ have a tinted background.
SpecificationHD2010UCDelta OHM (Senseca)Quest SoundPro SE/DLTSI Quest
StatusCurrentCurrent
Verified on
Standards and class
Standards (differs)IEC 61672-1:2013 class 1 or class 2, group X, IEC 61672-1:2002, IEC 60651:2001 and IEC 60804:2000 class 1 or 2EN/IEC 61672-1:2002 (class 1 or class 2 by model), ANSI S1.4 (R2006), ANSI S1.43 (R2007), EN/IEC 61260 (2001), ANSI S1.11 (R2009)
Pattern approvals (differs)I.N.RI.M. approval certificate no. 07-0124-02 (IEC 61672-1:2002)—
Measurement
Channels11
Measurement range (differs)30 dBA to 143 dB peak0 to 140 dB overall, over 8 individual ranges of 90 dB
Linearity range (differs)80 dB120 dB+ (A-weighted) total dynamic measurement range
Peak range (differs)up to 143 dB peakUp to 143 dB with BK4936 microphone
Frequency range (differs)—1/1 octave filters 16 Hz to 16 kHz; 1/3 octave filters 12.5 Hz to 20 kHz
Frequency weightingsA, C, ZA, C, Z
Time weightingsF, S, IF, S, I
Simultaneous weightings (differs)Yes—
Parameters (differs)SPL, Leq, LeqI, SEL, LEP,d, Lmax, Lmin, Lpeak, Dose, LN (three selectable L1 to L99; full L1 to L99 with HD2010.O2)SPL, Lmax, Lmin, Lpeak, Leq/Lavg, SEL, LN (L1 to L99 selectable), TWA, Dose, Ldn, CNEL
DosimetryYesYes
Exchange rates (differs)—3 dB, 4 dB, 5 dB, 6 dB
Analysis
Octave bands (differs)None1/1 and 1/3
Reverberation time (differs)—Yes
Speech intelligibility (differs)—STIPA (firmware OPTION-1, Acoustic Spectral Curves & Speech Intelligibility)
Audio recording (differs)No—
Logging and data
Shortest logging interval (differs)125 ms (one profile parameter, with HD2010.O2); report parameters 1 s to 1 h1 s (DL models; 11 selectable intervals to 60 min)
Memory (differs)8 MB internal flash, more than 500 records; SD card up to 2 GB via optional HD2010MC reader (requires HD2010.O2)SD card, 32 MB to 32 GB (card included)
Hardware
Microphone (differs)UC52/1 (class 1) or UC52 (class 2), 1/2" prepolarised free-field condenserClass 1: BK4936 1/2"; Class 2: QE7052 1/2"; preamp accepts 1/2" (13.2 mm) microphones; 1/4" and 1" optional
Detachable preamplifierYesYes
Outdoor microphone kit (differs)HDWME outdoor microphone protection; heated preamplifier HD2010PNE2W (option HD2010.OE)SP-OMS Outdoor Measuring System
Display (differs)Backlit graphic LCD, 128 x 64Transflective 128 x 64 dot-matrix LCD with backlight
Battery (differs)NiMH 4.8 V / 2.1 Ah rechargeable pack; external 5 to 24 V DC4 x AA alkaline (NiMH optional)
Battery life (differs)—10 h
Weight (differs)740 g540 g
Dimensions (mm) (differs)445 x 100 x 50 (with preamplifier)79 x 282 x 41 (3.1 x 11.1 x 1.6 in, with preamp and microphone)
IP rating (differs)IP64—
Operating temperature (°C)-10 to 50-10 to 50
Connectivity and software
Connectivity (differs)USB 1.1, RS-232C (300 to 115200 baud), AC line output (3.5 mm), DC output (2.5 mm), external power inputUSB 2.0 (mini-USB), RS-232 (10-pin auxiliary connector), AC/DC output (3.5 mm)
Software (differs)Noise Studio 1.0 Basic, NS1 Workers Protection, NS-ENS Environmental Noise Studio (requires HD2010.O2)Detection Management Software (DMS)

The instruments on their own