Spark 706RC vs TSI OmniWear Noise (OW-N-DVE)
Every specification Larson Davis and TSI Quest publish for these two instruments, side by side. Rows where the values differ are tinted. A blank cell means that manufacturer does not publish the figure, which is itself worth knowing. Nothing here is written by either manufacturer.
| Specification | Spark 706RCLarson Davis | TSI OmniWear Noise (OW-N-DVE)TSI Quest |
|---|---|---|
| Status | Discontinued | Current |
| Verified on | ||
| Standards and class | ||
| Standards (differs) | — | IEC 61252:1997 + A2:2017 (BS EN), ANSI S1.25:1991 (R2024) |
| Measurement | ||
| Channels | 1 | 1 |
| Measurement range (differs) | 40 to 143 dB rms in four ranges | 70 to 140.3 dB(A) RMS (linear operating range) |
| Linearity range (differs) | — | 70 to 140.3 dB(A) |
| Peak range (differs) | 80 to 146 dB in four ranges | 90.0 to 143.3 dB (C or Z weighted) |
| Frequency weightings (differs) | A, C | A, C, Z |
| Time weightings (differs) | F, S | F, S, I |
| Parameters (differs) | Leq, Lmax, Lpeak, Lmin, four time-weighted averages, Ln (1 to 99), Dose, projected Dose, SE (Pa2h), SEA | LXY, LXYmax, LXeq, LXpeak, Lavg, LXIeq, LTM3, LTM5, LAE, LAep,d, LEX,8h, TWA, Dose % (incl. projected), Pa2h, Pa2s, Motion %, time above threshold |
| Dosimetry | Yes | Yes |
| Exchange rates (differs) | 3 dB, 4 dB, 5 dB, 6 dB | 3 dB, 5 dB |
| Analysis | ||
| Octave bands (differs) | None | — |
| Reverberation time (differs) | No | — |
| Building acoustics (differs) | No | — |
| Audio recording (differs) | No | — |
| Logging and data | ||
| Shortest logging interval | 1 s | 1 s |
| Memory (differs) | 4 MB non-volatile | 64 runs maximum |
| Hardware | ||
| Microphone (differs) | 3/8" (10.6 mm) MPR001 integrated microphone/preamplifier on 1 m cable | MEMS microphone |
| Detachable preamplifier (differs) | No | — |
| Display (differs) | LCD with backlight (706RC only) | none (multifunction LED status indicator and one action button) |
| Battery (differs) | 2 x AA alkaline, NiCd or NiMH | Rechargeable Li-ion 350 mAh; wireless drop-in charger (4 h to 85 %, 6 h to 100 %) |
| Battery life (differs) | 100 h | 16 h |
| Weight (differs) | — | 36 g |
| Dimensions (mm) (differs) | — | 56 x 32 x 40 |
| IP rating (differs) | — | IP65 |
| Operating temperature (°C) (differs) | — | 0 to 40 (< +/-0.5 dB error); -10 to 50 (< +/-0.8 dB error) |
| Connectivity and software | ||
| Connectivity (differs) | infrared (DVX010 USB-to-IR module) | Bluetooth |
| Remote control (differs) | — | OmniWear Noise app (iOS, Android) |
| Software (differs) | Blaze | OmniWear Noise app |
The instruments on their own
- Spark 706RC — full record, sources and owner reviews
- TSI OmniWear Noise (OW-N-DVE) — full record, sources and owner reviews
Other comparisons
- Spark 706RC vs Quest Edge 4+ (EG4P)
- Spark 706RC vs Quest Edge 7 (EG7)
- Spark 706RC vs Quest Edge 8 (EG8)
- dBadge2 vs TSI OmniWear Noise (OW-N-DVE)